Knick Elektronische

Conductivity meter Portavo 902 Cond/904 Cond/904 X Cond | Type: 902 Cond

Article number: 9774710

€944.00*

Packaging unit: 1 piece

on order

Type
Product information "Conductivity meter Portavo 902 Cond/904 Cond/904 X Cond | Type: 902 Cond"
Portavo 902 Cond
  • The basic version for mobile conductivity measurement
  • Robust, intuitive portable device for routine daily measurements in laboratories and processes
  • A sensor quiver protects the sensor from drying out or being damaged in daily use
  • The high-performance polymer housing ensures low water absorption and high impact resistance
  • Over 1000 hours of measurement with a single set of batteries (4 x AA)
  • Memosens sensors and analog sensors can be used on the same device
  • The mineral glass display is perfectly readable even after years

Portavo 904 Cond
as Portavo 902 Cond, additionally with:
  • Li-ion battery
  • Data logger with 5,000 values
  • Micro USB port and Paraly SW 112 software

Portavo 904 X Cond
as Portavo 904 Cond, additionally with:
  • Conductivity measurement with ATEX approval for measurements up to Zone 0/1. Ideal for applications in the process industry. For the first time, Portavo enables direct on-site testing of process measuring points.
  • The new MemoLog function allows the user to log calibration data from various Memosens measuring points

Specifications
Measuring range (Accuracy):
0.001 µS/cm (c0.05 cm-1) 0.01 µS/cm (c=0.05 ... 0.2 cm-1) 0.1 µS/cm (c0.2 cm-1)
TDS: Salinity:
0 ... 1999 mg/l 0.0 ... 45.0 g/kg
Resistivity: Concentration:
00.00 ... 99.99 MO x cm 0.00 ... 9.99 (w/w) %
Conductivity input, analog:
Multi-contact for 2-/4-electrode sensors with integrated temp detector
Measuring ranges Sensor SE 202: Sensor SE 204:
0.01 ... 200 µS/cm 1 µS/cm ... 500 mS/cm
2-electrode sensors: 4-electrode sensors:
0.1 µS x c ... 200 mS x c 0.1 µS x c ... 1000 mS x c
Temperature input:
2 x 4 mm diam. for integrated or separate temperature detector
Messbereich mit NTC 30 kO: Messbereich mit Pt 1000:
-20 ... 120 °C -40 ... 250 °C
Conductivity input, Memosens:
M8 socket, 4 pins, for Memosens lab cable
Measuring range sensor SE 215 MS:
10 µS/cm ... 20 mS/cm
Conductivity input
Temperature compensation:
Linear 0 ... 20 %/K, reference temperature adjustable
nLF:
0 ... 120 °C
Power supply: Protection class acc. to DIN EN 60529:
4 x AA IP66/67 with pressure compensation
Type: 902 Cond